ARTHREX INC. SWVLK TENO BIO-COMP 7X 19.5MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
AR-1662BC-7 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/22/2015
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.This is a re-submission of an original submission per the fda because the original never made it into the maude database.Lot number was not provided so device history record review cannot be performed.The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The cause of the event could not be determined from the information available and without device evaluation.The most likely cause(s) of this event (screw coming loose) are improper bone preparation, implant over insertion in patients with insufficient quantity or quality of bone, and/or failure to follow the post-op rehab protocol.The directions for use (dfu-0087) states: post-operatively and until healing is complete, fixation provided by this device should be considered as temporary and may not withstand weight bearing or other unsupported stress.The fixation provided by this device should be protected.The post-operative regimen prescribed by the physician should be strictly followed to avoid adverse stresses applied to the device.The potential causes of this event are being communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the original surgery took place (b)(6) 2015.On (b)(6) 2015 when patient saw the surgeon, he noticed a lump where lhbt (long head of the biceps tendon) was.On (b)(6) 2015 a revision surgery was performed.The screw came loose and ended up resting in the deltoid.The bicep had integrated into the socket.The ar-1662bc-7 was removed.It was then secured with a button 10mm's below original drill hole.Explanted device was disposed of.Procedure was a proximal biceps.Male patient, age (b)(6).
|
| |
|
Search Alerts/Recalls
|
|
|