| Catalog Number |
816022100 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This is our initial report on this incident.
|
| |
|
Event or Problem Description
|
|
The customer reported a false negative result of a patient sample when tested with biotestcell-i11 plus on tango infinity.The customer stated that the specificity of the known antibody is anti-d due to the rh(d) prophylaxis.The customer returned neither the supposedly defective product nor the patient sample that had caused false negative test result.Testing of our retained product sample in our quality control laboratory is still ongoing.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of both allegedly defective lot.The affected tango infinity has been inspected by our field service engineers and found to be working within specifications.
|
| |
|
Event or Problem Description
|
|
The customer reported false negative results of a patient sample when tested with biotestcell-i11 plus on tango infinity.The customer stated that the specificity of the known antibody is an acquired anti-d due to the rh(d) prophylaxis.The patient sample was supposed to yield positive reactions with cell #1, #2, #3 and #4 of biotestcell-i11 plus.The customer returned neither the allegedly defective product sample for investigational testing nor the patient sample that had caused a false negative test result.Therefore our quality control laboratory tested their retention sample of biotestcell-i11 plus with controls and donor samples.All positive and negative reactions were correct.We did not observe any false negative reaction.Testing by our quality control laboratory confirmed the allegedly defective lot of biotestcell-i11 plus functions correctly.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of the allegedly defective lot.The affected tango infinity has been inspected by our field service engineers and found to be working within specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
This is our final report on this incident.
|
| |
|
Search Alerts/Recalls
|