• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ACC 2.0MM COCR CABLE W/CLAMP; CERCLAGE, FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ACC 2.0MM COCR CABLE W/CLAMP; CERCLAGE, FIXATION Back to Search Results
Catalog Number 71340007
Health Effect - Clinical Code No Code Available (3191)
Date of Event 11/20/2017
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed due to the patient's trochanteric region and greater trochanter was completely ununited.The cables and claw were removed.The trochanteric nonunion was fixed by reapproximating the greater trochanter in the appropriate position and placing a new cable claw construct in the appropriate position.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.The clinical/medical team concluded, unable to rule out the malpositioned prosthesis, leg-length discrepancy, ¿infection treatments¿, and continued motion along with the patient¿s comorbidities as likely contributing factors to the reported non-union of the trochanteric region/gt.However, without the supporting operative reports, lab/pathology results, imaging, and/or the analysis of the explanted components, the root cause cannot be determined and a causal relationship with the cables cannot be confirmed.The patient impact beyond the reported non-union, left hip revision left trochanteric/gt orif and expected post-op convalescence cannot be determined.No further medical assessment is warranted at this time.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed parts revealed no prior complaints for the listed batches.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACC 2.0MM COCR CABLE W/CLAMP
Common Device Name
CERCLAGE, FIXATION
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7808643
Report Number1020279-2018-01610
Device Sequence Number1573986
Product Code JDQ
UDI-Device Identifier03596010485731
UDI-Public03596010485731
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K031162
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2017
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71340007
Device Lot Number16KSM0185
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/31/2018
Supplement Date Received by Manufacturer07/31/2018
Initial Report FDA Received Date08/23/2018
Supplement Report FDA Received Date05/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-