Catalog Number 07028091190
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
The customer complained of discrepant high results for 1 patient sample tested for elecsys tsh on a cobas e 801 module.The patient was tested on the e801 module and the tsh result was 5.11 uiu/ml and was reported outside of the laboratory.The result was questioned by the doctor as the patient is taking l-thyroxine and repeat testing by another method was requested.On (b)(6) 2018, the sample was sent to an external laboratory using an unknown method and the tsh result was < 0.01 miu/ml.This result was what the doctor expected for this patient.There was no allegation that an adverse event occurred.The 801 module serial number was (b)(4).
 
Additional Manufacturer Narrative
Upon further investigation of the patient sample, an interferent against a component of the reagent was confirmed.This interference is covered in product labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TSH ASSAY
Common Device Name
RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key7808776
Report Number1823260-2018-02829
Device Sequence Number1573362
Product Code JLW
UDI-Device Identifier04015630939831
UDI-Public4015630939831
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K961491
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number07028091190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 08/03/2018
Supplement Date Received by Manufacturer08/03/2018
Initial Report FDA Received Date08/23/2018
Supplement Report FDA Received Date09/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
L-THYROXINE