| Catalog Number |
07028091190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
The customer complained of discrepant high results for 1 patient sample tested for elecsys tsh on a cobas e 801 module.The patient was tested on the e801 module and the tsh result was 5.11 uiu/ml and was reported outside of the laboratory.The result was questioned by the doctor as the patient is taking l-thyroxine and repeat testing by another method was requested.On (b)(6) 2018, the sample was sent to an external laboratory using an unknown method and the tsh result was < 0.01 miu/ml.This result was what the doctor expected for this patient.There was no allegation that an adverse event occurred.The 801 module serial number was (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Upon further investigation of the patient sample, an interferent against a component of the reagent was confirmed.This interference is covered in product labeling.
|
| |
|