MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number RONYX20022UX
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
During the procedure, an attempt was made to use one resolute onyx rx coronary drug eluting stent to treat a mildly tortuous and non-calcified lesion located in the mid ramus.There were no abnormalities reported in relation to the anatomy.There was no damage noted to packaging.There were no issues noted when removing the device from the hoop.The device was inspected with no issues noted.Negative prep was performed with no issues noted.The lesion was pre-dilated.The device did pass through a previously-deployed stent.Resistance was encountered when advancing the device.Excessive force was not used during delivery.It was reported that an attempt was made to pass the device distal to a previously deployed stent and stent deformation occurred in vivo.No patient injury reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Device evaluation: the stent was positioned on the balloon between the marker bands as per specifications.Misalignment was evident to the 1st distal stent wraps with struts raised.Deformation was evident to the distal tip.The 0.015 inch mandrel would not load through the guidewire lumen due to hardened blood on the catheter.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7809071
Report Number9612164-2018-02143
Device Sequence Number11091112
Product Code NIQ
UDI-Device Identifier00643169556669
UDI-Public00643169556669
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/27/2020
Device Model NumberRONYX20022UX
Device Lot Number0008972307
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2018
Initial Date Received by Manufacturer 08/06/2018
Supplement Date Received by Manufacturer09/20/2018
11/06/2018
Initial Report FDA Received Date08/23/2018
Supplement Report FDA Received Date10/18/2018
11/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age56 YR
Patient Weight87
-
-