• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA TIBIAL INSERT FIXED SPHERE FLEX SIZE 2/13 MM R

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA TIBIAL INSERT FIXED SPHERE FLEX SIZE 2/13 MM R Back to Search Results
Catalog Number 02.12.0211FR
Health Effect - Clinical Code No Code Available (3191)
Date of Event 07/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As reported by the field, no torque limiter used in primary surgery.Batch review performed on 23 august 2018.Lot.164442: (b)(4) items manufactured and released on 19 october 2016.Expiration date: 2021-09-19 no anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Clinical evaluation performed on 23 august 2018.Sixteen (16) months after primary tka the insert fixation screw got loose in the joint.It was immediately removed with an arthroscopic operation.No consequence should be expected for the absence of the screw in the future life of the implant.During arthroscopic surgery, implant surfaces can be visualized and checked for damages.The cause for self-unscrewing of the insert screw is not known: one possibility is insufficient tightening torque, but other unknown conditions may also play a role.Visual inspection of the explanted tibial insert secure fixation screw on 23 august 2018 by r and d product manager: loosened insert safety screw occurred about 1 year and 4 months after primary surgery and removed by arthroscopy.The head and the threaded part of the screw looks partially damaged.They have most likely been plastically pressed once, backed out from the baseplate, were interposed between the femoral component and the tibia insert and underwent to the body load.Tibial insert, tibial baseplate and femoral components have not been explanted, so we can suppose that they have been found in good condition.The most-likely cause for self-unscrewing of the screw after few months from implantation, was insufficient tightening torque.Using a screwdriver provided with a torque limitation (ref.02.07.10.4577; torque limiter screwdriver 3.5 nm) would prevent this event.Torque limiter screwdriver is already available and its use is mandatory.But other unknown factors might also play a role in the event.No further considerations can be done basing on the available information.
 
Event or Problem Description
Patient presented 16 months post surgery right total knee replacement sphere knee.Tibial insert screw backout noted on x ray.Patient had screw removed arthroscopically under a general anaesthetic.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL INSERT FIXED SPHERE FLEX SIZE 2/13 MM R
Common Device Name
TIBIAL INSERT FIXED
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
MDR Report Key7809075
Report Number3005180920-2018-00631
Device Sequence Number19199716
Product Code JWH
UDI-Device Identifier07630030862540
UDI-Public07630030862540
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K140826
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/19/2021
Device Catalogue Number02.12.0211FR
Device Lot Number164442
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2018
Initial Report FDA Received Date08/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient Weight62
-
-