• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC HIWIRE NITINOL HYDROPHILIC WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC HIWIRE NITINOL HYDROPHILIC WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Catalog Number HWS-038150
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, during an unknown procedure, the teflon on the hiwire nitinol hydrophilic wire guide split/peeled.Additional information has been requested regarding the patient and the event.At the time of this report, no further information has been provided.No adverse events have been reported as a result of the alleged malfunction.
 
Additional Manufacturer Narrative
Additional information received 24aug2018.Method codes: 4114, device not returned.Conclusion codes: 4135, cause not established.Investigation evaluation: the complaint device was not returned for an evaluation and no photograph or imaging was provided.Without the complaint device, a device failure analysis was unable to be performed.A document based investigation was conducted including a review of complaint history, the device history record, and quality control data.The device history record was reviewed noted there are no conformances noted.A review of complaint history records revealed there have been no other similar complaints associated with lot number 10984545.The cause of the complaint was not able to be established.Per the quality engineering risk assessment, no additional risk mitigating activity is required at this time.Cook medical has notified the appropriate personnel and will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Additional information received.As reported, while performing a ureteroscopy the surgeon noticed that the ¿teflon¿ on the guidewire kind of split/peeled a little.No unintended section of the device remained inside the patient's body.There were no additional procedures required due to this occurrence and there were no adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIWIRE NITINOL HYDROPHILIC WIRE GUIDE
Common Device Name
OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7809360
Report Number1820334-2018-02556
Device Sequence Number1293522
Product Code OCY
UDI-Device Identifier00827002304703
UDI-Public(01)00827002304703(17)210326(10)10984545
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K082536
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/26/2021
Device Catalogue NumberHWS-038150
Device Lot Number10984545
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/10/2018
Supplement Date Received by Manufacturer10/08/2018
Initial Report FDA Received Date08/23/2018
Supplement Report FDA Received Date10/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ASKU; ASKU
-
-