| Catalog Number |
03260763160 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer complained of false positive nitrite results for over 40 patient urine samples on a urisys 1100 analyzer serial number (b)(4) since (b)(6) 2018.The specific results were asked for but not provided.There were erroneous results reported outside of the laboratory.There was no allegation of an adverse event.No treatment or actions were taken based on the results.The customer stated that they dipped a test strip in water and the nitrite result was a falsely positive.The customer stated that the nitrites portion of a test strip that had not been used appeared "pinkish".The customer stated that the calibration was acceptable.The suspect product was requested to be returned for investigation.The retention material of lot 27638300 was measured on an iu urisys 1800 with native urine and a nitrite dilution series.The results fulfilled the requirements.No false positive results were observed.
|
| |
|
Additional Manufacturer Narrative
|
|
Medwatch fields device available for evaluation? and device evaluated by mfr? were updated.A vial of test strips was returned for investigation.The investigation determined that the nitrite test pads of all the returned test strips were discolored.There were no discolored test pads in the retention material.The error was consistent with mishandling.In relation to strip handling the package insert states "in order to avoid exposure to moisture, the vial must be closed immediately after removal of a strip, using the original stopper, which contains a drying agent.".
|
| |
|
Search Alerts/Recalls
|