TRIVIDIA HEALTH, INC. TRUETRACK; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUETRACK
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/31/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product not yet returned for evaluation.(b)(4).Manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Event or Problem Description
Consumer reported complaint for high blood glucose test results.Friend is calling on behalf of the customer.The customer is concerned with tests results from results obtained of 392 and 254 mg/dl.The customer's expected fasting blood glucose test result range is 90 - 180 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The customer declined to perform a back to back blood test during the call.The product is stored according to specification in the living room.The test strip lot manufacturer's expiration date is 07/31/2020 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history (date/time not set): result 1 :195mg/dl date: (b)(6) 2018 time:12:40pm not fasting; result 2 :192mg/dl date: (b)(6) 2018 time:12:40pm not fasting; result 3 :181mg/dl date: (b)(6) 2018 time:12:48pm not fasting; result 4 :392mg/dl date: (b)(6) 2018 time:12:40pm fasting; result 5 :254mg/dl date: (b)(6) 2018 time:07:15pm fasting.
 
Event or Problem Description
Consumer reported complaint for high blood glucose test results.Friend is calling on behalf of the customer.The customer is concerned with tests results from results obtained of 392 and 254 mg/dl.The customer's expected fasting blood glucose test result range is 90 - 180 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The customer declined to perform a back to back blood test during the call.The product is stored according to specification in the living room.The test strip lot manufacturer's expiration date is 07/31/2020 and open vial date is 07/11/2018.The meter memory was reviewed for previous test result history (date/time not set): result 1 :195mg/dl, date: (b)(6) 2018, time:12:40pm, not fasting; result 2 :192mg/dl, date: (b)(6) 2018, time:12:40pm, not fasting; result 3 :181mg/dl, date: (b)(6) 2018, time:12:48pm, not fasting; result 4 :392mg/dl, date: (b)(6) 2018, time:12:40pm, fasting; result 5 :254mg/dl, date: (b)(6) 2018, time:07:15pm, fasting.
 
Additional Manufacturer Narrative
(manufacturer narrative = t, corrected data = f) internal report # (b)(4) product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi#: (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.Correction: add on h10 the mlurc for the complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUETRACK
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7810147
Report Number1000113657-2018-00888
Device Sequence Number13846281
Product Code NBW
UDI-Device Identifier00021292001018
UDI-Public(01)00021292001018
Combination Product (Y/N)N
Initial Reporter StateNY
PMA/510(K) Number
K032657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 08/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/31/2020
Device Model NumberTRUETRACK
Device Catalogue NumberA4H01-81
Device Lot NumberRV5239
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Initial Date Received by Manufacturer 07/31/2018
Supplement Date Received by Manufacturer07/31/2018
Initial Report FDA Received Date08/23/2018
Supplement Report FDA Received Date08/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SECOND THERAPY; SECOND THERAPY
-
-