• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK HEALIX ADVANCE 5.5 MM BR MULTPACK - 2 SUTURE (QTY 5); SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK HEALIX ADVANCE 5.5 MM BR MULTPACK - 2 SUTURE (QTY 5); SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Catalog Number 222323
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/04/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device is not being returned, therefore is unavailable for a physical evaluation.A batch record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy synthes mitek complaints system revealed one similar complaint for this lot released to distribution.We cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.This file will remain receptive to any potential forthcoming information received that is pertinent and germane to this issue.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event or Problem Description
It was reported by the sales rep that during a rotator cuff repair, it was observed that the blue suture on the customer's healix advance 5.5mm br anchor with two sutures broke near the end of the strand when the surgeon had just started to lightly tension the knot.The surgeon completed the procedure by securing the anchor with the remaining part of the blue suture to the violet suture with no patient consequences or delays.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALIX ADVANCE 5.5 MM BR MULTPACK - 2 SUTURE (QTY 5)
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK
325 paramount drive
raynham MA 02767
MDR Report Key7810286
Report Number1221934-2018-53961
Device Sequence Number1469045
Product Code MAI
UDI-Device Identifier10886705021512
UDI-Public10886705021512
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K120078
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/11/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2016
Device Catalogue Number222323
Device Lot Number3748786
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/11/2015
Initial Report FDA Received Date08/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-