| Catalog Number |
06407137190 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
The customer stated that they received erroneous results for 11 patient samples tested for crej2 creatinine jaffé gen.2 on a cobas 8000 c 702 module that were tested in sequence.The initial erroneous results were reported outside of the laboratory.The samples were repeated and the repeat results were not reported outside of the laboratory as the customer believed there was no clinical difference between the initial and repeat results.No adverse events were alleged to have occurred with the patients.The serial number of the c 702 analyzer is (b)(4).Upon visual inspection, all samples were clear and did not contain fibrin.The number of tests left in the reagent pack was low, so the field service representative changed the reagent pack, calibrated, and then ran controls.The customer measured controls every 4 hours to monitor the issue and no other issues have were noted.Calibration and controls were within expected ranges, however the customer repeated controls after the low patient values were noticed and controls had lower values similar to the affected patient values.Upon review of the alarm trace, some low water bath volume alarms were observed.There were also alarms indicating that the water reservoir level was too low and that there was an issue with a pump.
|
| |
|
Additional Manufacturer Narrative
|
|
Co2 at normal concentrations does not interfere with the assay.High concentrations of ambient co2 can cause a reduction in the ph of the reagent, which can lead to lower results.Product labeling requires calibration of this assay at least every 24 hours that a reagent pack has been onboard the analyzer.In this case the customer had not calibrated for greater than 24 hours.If the customer had calibrated per the package insert the calibration would have compensated for any ph drift.
|
| |
|
Additional Manufacturer Narrative
|
|
Precision studies were performed on the analyzer and based on the results, a hardware issue could be excluded.Since there was a constant lower drift in patient results, the observed issue is consistent with contamination by high environmental co2 concentrations.
|
| |
|
Search Alerts/Recalls
|