If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device was released and evaluated.Visual inspection confirms the proximal end of the screw is broken.It is unknown if the insertion was off axis, if the procedure was a btb or st.In case of a btb procedure and especially in hard bone, notching and tapping is required before inserting the screw.It was reported that the channel was 7mm and a 8x30mm screw was used.No further technique details were provided that could help determine a root cause.Furthermore, no lot numbers were supplied for the applier or the needles which precludes conducting a batch history review or a lot specific search in the complaints handling system.At this point in time, no further action is warranted.However, mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed from the (b)(4) complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|