Catalog Number 3548212
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/31/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A device history record review is in progress.Once completed, a supplemental report will be submitted.(b)(4).
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure with mentor cpx 4 breast tissue expander 350cc saline breast implants that intraoperatively when surgeon tried to fill with saline it leaked.He removed the tissue expander and tested it outside the patient and confirmed it was leaking.They opened another tissue expander and inserted it and filled it with saline and everything went well.No patient consequences reported.
 
Additional Manufacturer Narrative
On (b)(6) 2018, the mentor failure analysis lab has received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.(b)(4).
 
Additional Manufacturer Narrative
On 8/23/2018, the device history record (dhr) was reviewed, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.(b)(4).
 
Additional Manufacturer Narrative
Additional information received on 8/27/2018, indicated that the date of implantation and explant was on (b)(6) 2018.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Device evaluation summary: upon receipt by mentor, the device was received with white fluid.The device appeared intact.Because the authorization for return and examination of medical device form was not signed and/or returned to mentor, pe was precluded from further evaluating the device.No anomalies were observed.Complaint was not confirmed.Mentor products are 100% visually inspected prior to release in addition to thorough in-process testing during several stages of the manufacturing process.A root cause failure analysis will not be conducted since the investigation of the reported failure could not confirm that an actual failure of the device to conform to expected performance had occurred.All the implants are 100% inspected before leaving the facility.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR CPX 4 BREAST TISSUE EXPANDER
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key7811068
Report Number1645337-2018-05146
Device Sequence Number19210084
Product Code LCJ
UDI-Device Identifier00081317006534
UDI-Public00081317006534
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 07/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/20/2022
Device Catalogue Number3548212
Device Lot Number7566511
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2018
Initial Date Received by Manufacturer 07/30/2018
Supplement Date Received by Manufacturer07/30/2018
07/30/2018
09/13/2018
10/09/2018
Initial Report FDA Received Date08/23/2018
Supplement Report FDA Received Date08/24/2018
08/28/2018
09/17/2018
10/09/2018
Patient Sequence Number1
-
-