• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD SOFSILK; SUTURE, NONABSORBABLE, SILK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD SOFSILK; SUTURE, NONABSORBABLE, SILK Back to Search Results
Model Number GSJ-34-M
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, intra-operatively, the nurse found the tip of the needle crushed.This device was not used.The event occurred during the procedure, but the product was not used for patient.The procedure was completed with another device.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection noted that all six needles were attached to the foam of the retainer and the sutures remained inside the retainer.The tips of the needles showed no abnormalities.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Additionally, the investigation detected a secondary condition of incorrect needle/suture quantity that has no relationship to the reported condition.This failure determined to be a result of a manufacturing activity, manpower error.Improvements have been initiated to mitigate this condition.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFSILK
Common Device Name
SUTURE, NONABSORBABLE, SILK
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7811257
Report Number9612501-2018-01509
Device Sequence Number1116770
Product Code GAP
UDI-Device Identifier20884521085319
UDI-Public20884521085319
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K981128
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2023
Device Model NumberGSJ-34-M
Device Catalogue NumberGSJ-34-M
Device Lot NumberD8A1136X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/24/2018
Initial Date Received by Manufacturer 08/06/2018
Supplement Date Received by Manufacturer02/15/2019
04/04/2019
Initial Report FDA Received Date08/23/2018
Supplement Report FDA Received Date03/11/2019
05/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-