BARD ACCESS SYSTEMS POWERGLIDE MIDLINE CATHETER, 18G 10CM; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
|
Back to Search Results |
|
| Model Number |
M118100 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of recn1704 showed one other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported via sales rep from facility that "mid line was being placed in the cephalic vein, wire would not deploy due to resistance met , wire did not lock into place due to resistance met immediately, wire was pulled back and catheter was repositioned via ultra sound guidance, wire deployed and catheter advanced without resistance, catheter would not flush and was pulled out, distal tip of catheter kinked approximately 1 cm from the end of the catheter.".
|
| |
|
Additional Manufacturer Narrative
|
|
The initial complaint appeared to be a reportable occurrence.Once the sample was returned and evaluated it was determined that a reportable event had not occurred.The returned sample exhibited no evidence of the alleged event.The event is unconfirmed per the investigation results.Since no device malfunction occurred and there is no allegation of a serious injury to a patent, user, or other person, this event is deemed not reportable per 21 cfr part 803.
|
| |
|
Event or Problem Description
|
|
It was reported via sales rep from facility that "mid line was being placed in the cephalic vein, wire would not deploy due to resistance met , wire did not lock into place due to resistance met immediately, wire was pulled back and catheter was repositioned via ultra sound guidance, wire deployed and catheter advanced without resistance, catheter would not flush and was pulled out, distal tip of catheter kinked approximately 1 cm from the end of the catheter.".
|
| |
|