• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. DISPOSABLE NERVE STIMULATOR - VARI-STIM® III HAND-HELD; STIMULATOR, NERVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. DISPOSABLE NERVE STIMULATOR - VARI-STIM® III HAND-HELD; STIMULATOR, NERVE Back to Search Results
Model Number 8562010
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No evaluation has been completed at the time of this report as the device has not been returned.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A healthcare professional reported that the varistim did not work.It was the initial use of the device.The user opened another varistim and it worked right away.There was no injury; no patient impact other than a short delay to open another package of varistim.
 
Additional Manufacturer Narrative
One opened sample of part number 8562010, from lot number 0214664249 was received for analysis.Visually, there was no damage or anomalies in the construction of the device that would have resulted in the reported event.The device was improperly aligned with the 2.0 ma position when received which would not have provide stimulating current.Note ¿ the device has 3 positions and when the marker is aligned to the desired output it clicks in place.The switch was rotated to align with the 2.0 ma position for the analysis and functionally tested per the ifu; the device would light which indicates it was delivering stimulating current.The device was then functionally tested in the other two positions with no issues or intermittent behavior while manipulating the switch and tip within the positions.The device was electrically tested with no out of specification condition identified: ¿ position 0.5ma requirement is 0.5ma +/- 0.1ma (0.4ma - 0.6ma), and measures 0.51ma.¿ position 1.0ma requirement is 1.0ma +/- 0.2ma (0.8ma - 1.2ma), and measures 1.05ma.¿ position 2.0ma requirement is 2.0ma +/- 0.4ma (1.6ma - 2.4ma), and measures 2.08ma.There was no evidence of improper manufacturing and no fault was found as it relates to the complaint; therefore, manufacturing has been ruled out as a likely cause.The ifu indicates: ¿select output amperage by rotating the probe index until marker is aligned with desired output¿; confirm functionality by touching the probe tip and needle electrode; and provides stimulation technique.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE NERVE STIMULATOR - VARI-STIM® III HAND-HELD
Common Device Name
STIMULATOR, NERVE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key7812284
Report Number1045254-2018-00385
Device Sequence Number1738900
Product Code ETN
UDI-Device Identifier20643169341818
UDI-Public20643169341818
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K905035
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/04/2020
Device Model Number8562010
Device Catalogue Number8562010
Device Lot Number0214664249
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/30/2018
Supplement Date Received by Manufacturer09/20/2018
Initial Report FDA Received Date08/24/2018
Supplement Report FDA Received Date10/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-