| Catalog Number |
636008P |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Blood Loss (2597)
|
| Date of Event |
07/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A review of similar events was performed for all gelsoft branded devices that have been reported to have leaked and that were impregnated with russelot or gelita gelatin.The incident rate for reported leakage events was confirmed as (b)(4) for products impregnated with both gelatin types.This is the first compliant for leakage for the grafts impregnated with gelita gelatin (occurrence rate of (b)(4) complaints v sales).The review included all similar events reported to vascutek between jan2013 - jul2018 inclusive.No negative trend was identified during the review.Vascutek awaiting further information from site.
|
| |
|
Event or Problem Description
|
|
Vascutek ltd.Was notified of an event that occurred in (b)(6), the event was described as follows: during femoro-popliteal bypass surgery, the prosthesis gelsoft plus presented many defects once it was in load (after completion of a proximal anastomosis).Many small holes have been seen all along the graft, inducing massive blood loss greater than 900cc for the patient, filling made with 1250cc of crystalloid and 2 pellets of gr for transfusion.Patient is doing well.There was no post-operative leakage observed.
|
| |
|
Additional Manufacturer Narrative
|
|
Cause not established.The root cause of the event could not be identified as device was not available for further analysis and additional information was not received from the customer.Three attempts were made to obtain additional information about the event.Further action is not planned; however, the issue will be tracked and trended as part of the on-going complaints trending and reporting process and if an adverse trend develops action may be taken at that time.(b)(4) consider this case closed.
|
| |
|
Event or Problem Description
|
|
This report is being submitted as follow up # 1 for mfr.Report # 9612515-2018-00012 to provide closure information for the case.
|
| |
|