Model Number SIGLU45A
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during robotic liver resection, while stapling the liver, after a few firings the device would not open once placed into the patient.To complete the case the surgeon needed to push the cartridge side against tissue to make it open.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Microscopic evaluation noted that the loading unit had damage to the cutting edge of the knife blade.Functionally the loading unit was loaded with a representative cartridge and was applied to test media.All staples were placed, and test media was cleanly transected.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Additionally, the investigation detected a secondary condition of obstruction and knife blade damage that has no relationship to the reported condition.Replication of the damaged knife blade may occur when an obstacle has been incorporated in the jaws during application.The information booklet which accompanies each product shipment cautions the user; ensure that no obstructions (such as clips) are incorporated in the instrument jaws.Firing over an obstruction may result in incomplete cutting action and/or improperly formed staples.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGNIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key7812688
Report Number1219930-2018-04655
Device Sequence Number1746906
Product Code GDW
UDI-Device Identifier20884521543499
UDI-Public20884521543499
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K160176
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model NumberSIGLU45A
Device Catalogue NumberSIGLU45A
Device Lot NumberN7K0672X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2018
Initial Date Received by Manufacturer 08/07/2018
Supplement Date Received by Manufacturer01/16/2019
Initial Report FDA Received Date08/24/2018
Supplement Report FDA Received Date02/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age66 YR
Patient Weight122
-
-