| Catalog Number |
210110 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
Leg kept falling to the side while locked into place.Case type: tka.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported issue: it was reported that the leg kept falling to the side.Product history review: review of the device history records indicate (b)(4) devices were manufactured and accepted into final stock on 02/02/2018.No non-conformances were identified during inspection.Complaint history review a review of complaints related to p/n 210110, lot 19360717 shows no additional complaint(s) related to the failure in this investigation.Material analysis material analysis was not completed as no material failure was alleged.Functional inspection: functional inspection shows that the ball is loose in the sled.The metal retainer rings that tighten over the ball are loose inside their housing.Conclusion: the failure mode was confirmed.Corrective action/preventive action: a review of stryker¿s nc/capa database indicated there has been no nc or capa associated with the product and failure mode reported in this event.
|
| |
|
Event or Problem Description
|
|
Leg kept falling to the side while locked into place.Case type: tka.
|
| |
|