• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. SLED ASSEMBLY; STEREOTAXIC DEVICE, ROBOTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. SLED ASSEMBLY; STEREOTAXIC DEVICE, ROBOTICS Back to Search Results
Catalog Number 210110
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
 
Event or Problem Description
Leg kept falling to the side while locked into place.Case type: tka.
 
Additional Manufacturer Narrative
Reported issue: it was reported that the leg kept falling to the side.Product history review: review of the device history records indicate (b)(4) devices were manufactured and accepted into final stock on 02/02/2018.No non-conformances were identified during inspection.Complaint history review a review of complaints related to p/n 210110, lot 19360717 shows no additional complaint(s) related to the failure in this investigation.Material analysis material analysis was not completed as no material failure was alleged.Functional inspection: functional inspection shows that the ball is loose in the sled.The metal retainer rings that tighten over the ball are loose inside their housing.Conclusion: the failure mode was confirmed.Corrective action/preventive action: a review of stryker¿s nc/capa database indicated there has been no nc or capa associated with the product and failure mode reported in this event.
 
Event or Problem Description
Leg kept falling to the side while locked into place.Case type: tka.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLED ASSEMBLY
Common Device Name
STEREOTAXIC DEVICE, ROBOTICS
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key7812814
Report Number3005985723-2018-00513
Device Sequence Number16376263
Product Code OLO
UDI-Device Identifier00848486031848
UDI-Public00848486031848
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K172219
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number210110
Device Lot Number19360717
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 08/20/2018
Supplement Date Received by Manufacturer10/09/2018
Initial Report FDA Received Date08/24/2018
Supplement Report FDA Received Date10/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-