• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 1.5MM DRILL GUIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 1.5MM DRILL GUIDE Back to Search Results
Model Number 312.155
Medical Device Problem Codes Break (1069); Material Deformation (2976)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There was no known reported patient involvement associated with the complained event.Unknown when device malfunctioned.Device is an instrument and is not implanted/explanted.Reporter is synthes employee.A device history record (dhr) review was conducted for part # 312.155, synthes lot # 6005159, supplier lot # na: release to warehouse date: 31dec2008, manufactured by synthes (b)(4): no non-conformance reports (ncr¿s) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.A product investigation was conducted.The instrument was received with the reported condition of ¿visual deformed/bent: nrm & broken (2+ pieces) pre or post operatively/step unknown: nrm¿.The following investigation flow were performed: damage: visual (appearance not as expected).It was visually observed that the drill guide is not broken instead the drill guide portion of the instrument is bent inward at the juncture between the guide and handle.The balance of the device is in fair condition.The received condition does agree with the complaint description for bent, however not for broken.The complaint is confirmed.Dimensional inspection: drill guide length and shaft diameter were measured and found to be conforming per relevant drawing.A review of the current design drawing and the drawing revision at the time of manufacture was performed and no issues were identified.Based on the available information it is not possible to determine a definitive root cause for the complaint condition a possible cause could be rough handling during the device¿s 7+ year lifespan has resulted in the complaint condition.During the investigation, no product design or manufacturing issues were observed that may have contributed to the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during an inspection of sets, the loaners department found that the one (1) drill guide and two (2) matrixmidface drill bits were broken, bent and non-repairable.There was no patient involvement.This is report 3 of 3 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5MM DRILL GUIDE
Common Device Name
GUIDE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7813355
Report Number2939274-2018-53455
Device Sequence Number19340254
Product Code FZX
UDI-Device Identifier10887587036489
UDI-Public(01)10887587036489
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number312.155
Device Catalogue Number312.155
Device Lot Number6005159
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/23/2018
Initial Report FDA Received Date08/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-