Model Number MMT-751NAP |
Device Problems
No Display/Image (1183); Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/26/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
The customer reported via phone call that the insulin pump had a blank display and audio/vibration anomaly.The customer¿s blood glucose level was 138 mg/dl at the time of the incident.Customer woke up this morning and the insulin pump display was blank and had a high pitch noise.The customer was assisted with troubleshooting and the display did not return.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be return for analysis.
|
|
Manufacturer Narrative
|
Insulin pump passed displacement test, self test, off no power test and all operating currents tested within the spec range.Insulin pump was monitored and tested with no blank display and audio anomaly noted.(b)(4).
|
|