DEPUY MITEK LLC US INTRAFIX TIBSHEATHINSERT FIXHNDLE *EA; ORTHOPAEDIC IMPLANT INSERTER/EXTRACTOR Back to Search Results
Catalog Number 254616
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/27/2013
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement: the complaint device was received and evaluated.Visual inspection confirms the distal tip is completely broken off.The returned broken pieces were examined and can be concluded that no pieces are missing.All broken pieces were returned.The storage/ cleaning details of this device is unknown and therefore a root cause cannot be determined.Furthermore, no lot numbers were supplied which precludes conducting a batch history review or a lot specific search in the complaints handling system.At this point in time, no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event or Problem Description
It was reported that upon return of a loaner kit, the tip of one (1) intrafix tibial sheath inserter, fixed handle was found broken.This case has been reported by the loan kit technician during loan kit inspection.The instrument had further been assessed by depuy engineering.No further information can be obtained as the case was not reported by a customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
ORTHOPAEDIC IMPLANT INSERTER/EXTRACTOR
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7813767
Report Number1221934-2018-51354
Device Sequence Number13879812
Product Code GEF
UDI-Device Identifier10886705013142
UDI-Public10886705013142
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/27/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254616
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2014
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/27/2013
Initial Report FDA Received Date08/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-