Model Number E SERIES
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event or Problem Description
Complainant alleged that during functional testing, the device would not power up.Complainant did not indicate that there was any patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
The device was returned to zoll medical canada's service department.The reported malfunction was observed and attributed to the lithium battery on the system board.The device was recertified and returned to the customer.Zoll medical corporation recommends replacement of the lithium battery on the system board every 5 years.This device is approximately 11 years old from date of manufacture.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E SERIES DEFIBRILLATOR
Common Device Name
E SERIES
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key7813769
Report Number1220908-2018-02417
Device Sequence Number1941071
Product Code MKJ
UDI-Device Identifier00847946008840
UDI-Public00847946008840
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K042007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberE SERIES
Device Catalogue NumberE SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/03/2018
Supplement Date Received by Manufacturer08/03/2018
Initial Report FDA Received Date08/24/2018
Supplement Report FDA Received Date09/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-