TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUEMETRIX
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Returned meter evaluated with no defect found.Returned test strips evaluated with defect found.Qc investigation on 8/9/2018 resulted in defect found; returned test strips produce high readings and the event was determined to be reportable.R&d final root cause: most likely underlying root cause of high readings are due to improper storage: vial left open for an extended period of time and returned vial cap was broken off and missing.Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Event or Problem Description
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 140, 125, 215, 269 and 323 mg/dl.Customer did not know if the results were fasting or non-fasting.Customer did not know his expected blood glucose test result range.During the call on (b)(6) 2018, the customer felt well and did not report any symptoms.Medical intervention is not reported at the time of the call on (b)(6) 2018 as a result of the meter's readings.Blood test performed non-fasting during call on (b)(6) 2018 produced result of 318 mg/dl using truemetrix meter.The product is not stored according to specification and is stored in the kitchen.The test strip lot manufacturer's expiration date is 10/27/2019 and open vial date is one month.During the call on (b)(6) 2018, the customer mixed the test strips with another brand of strips.Adverse event not reported at time of call on (b)(6) 2018.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7813867
Report Number1000113657-2018-00892
Device Sequence Number19329415
Product Code NBW
UDI-Device Identifier00021292007041
UDI-Public(01)00021292007041
Combination Product (Y/N)N
Initial Reporter StateMI
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 08/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/27/2019
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMV2872
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2018
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/09/2018
Initial Report FDA Received Date08/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-