• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET HELICAL NANO 1MM X 3CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET HELICAL NANO 1MM X 3CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Catalog Number M0035431030
Medical Device Problem Codes Break (1069); Unintended Movement (3026); Migration (4003)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/07/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The subject device remains implanted.
 
Event or Problem Description
It was reported that during the coil embolization procedure, the coil broke, protruded and migrated out of the aneurysm.The physician used a guidewire to push the coil back into the aneurysm and procedure was completed.There were no reported clinical consequences to the patient.
 
Additional Manufacturer Narrative
The device history record review confirms that the device met all material, assembly and performance specifications.The subject device is not available; therefore, functional testing as well as physical analysis cannot be performed.The cause of the reported issue could not be definitively determined.
 
Event or Problem Description
It was reported that during the coil embolization procedure, the coil broke, protruded and migrated out of the aneursym.The physician used a guidewire to push the coil back into the aneurysm and procedure was completed.There were no reported clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET HELICAL NANO 1MM X 3CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key7813931
Report Number3008881809-2018-00363
Device Sequence Number169795
Product Code HCG
UDI-Device Identifier04546540698018
UDI-Public04546540698018
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/20/2021
Device Catalogue NumberM0035431030
Device Lot Number20389185
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/08/2018
Supplement Date Received by Manufacturer09/21/2018
Initial Report FDA Received Date08/24/2018
Supplement Report FDA Received Date09/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ECHELON MICROCATHETER(MEDTRONIC)
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age72 YR
-
-