COOK INC MULTI-SIDEPORT DUAL CHECK VALVE CATHETER INFUSION SET; DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Catalog Number MCIS-5.0-35-130-15.0
Medical Device Problem Code Unraveled Material (1664)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: 5fr ((b)(4)) terumo introducer, terumo stiff guidewire, amplatz superstiff guidewire.Pma/510(k) number = pre-amendment.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, the multi-sideport dual check valve catheter infusion set displayed "an excessive elastic elongation with risk of rupture" upon removal from the patient's anatomy.During the procedure to treat a right femoral-popliteal obstruction with local fibrinolytic treatment, there was difficulty in advancing the complaint device from the humeral access to the right femoral-popliteal obstruction.Per the complainant, the patient had diffuse aortic and iliac calcifications but no tortuosity of the vasculature.After removal of the complaint device, the procedure was completed using a second catheter of the same type.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.According to the complainant, the complaint device will not be returned to the manufacturer for investigation.
 
Additional Manufacturer Narrative
Investigation-evaluation: a review of the complaint history, drawings, device history record, documentation, manufacturing instructions, and quality control was conducted during the investigation.The complaint device was not returned; therefore, no physical examination could be performed.However, a document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Review of the device history record of the finished product shows no nonconforming events that could contribute to this failure mode.A complaint history search revealed that there was one other reported complaints for this lot number.Based on the information provided, no product returned, and the results of our investigation, a definitive root cause could not be determined.Per the [quality engineering] risk assessment, no further action is required.Monitoring will continue to be performed for similar complaints.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTI-SIDEPORT DUAL CHECK VALVE CATHETER INFUSION SET
Common Device Name
DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7814063
Report Number1820334-2018-02499
Device Sequence Number1707428
Product Code DQO
UDI-Device Identifier00827002064256
UDI-Public(01)00827002064256(17)200918(10)8154876
Combination Product (Y/N)N
Initial Reporter CountryIT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/18/2020
Device Catalogue NumberMCIS-5.0-35-130-15.0
Device Lot Number8154876
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/08/2018
Supplement Date Received by Manufacturer10/22/2018
Initial Report FDA Received Date08/24/2018
Supplement Report FDA Received Date10/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age73 YR
Patient Weight65
-
-