• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B.BRAUN SURGICAL SA NOVOSYN VIOLET 0 (3.5) 90CM HR37S (M); SUTURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B.BRAUN SURGICAL SA NOVOSYN VIOLET 0 (3.5) 90CM HR37S (M); SUTURE Back to Search Results
Model Number C0068956
Device Problems Device Misassembled During Manufacturing /Shipping (2912); Packaging Problem (3007); Inaccurate Information (4051)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/25/2018
Event Type  malfunction  
Event Description
Country of complaint: (b)(6).Wrong product inside box.
 
Manufacturer Narrative
Samples received: none but pictures received.Analysis and results: there are no previous complaints of this code batch of which we manufactured and distributed (b)(4) units in the market.There are no units in stock in b.Braun surgical's warehouse.Regarding to the product found inside the box (b)(4), we manufactured and mostly distributed (b)(4) units into the market.There are 24 units in stock in b.Braun surgical's warehouse.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfill usp/ep and b.Braun surgical requirements.Reviewed the batch manufacturing record of the product found inside the box, there is an internal non conformity related to a labeling issue.Both products were packed very close in time in the manufacturing area.We assume that the clean line was not correctly performed and some units of the previous order (c1049619, batch 118113) were packed with the product c0068956 and batch 118114.Final conclusion: taking into account that the results of the analysis show that the product did not fulfil the specifications of the european pharmacopoeia/b.Braun surgical specifications, we conclude that the complaint is confirmed by evidence of the pictures received.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVOSYN VIOLET 0 (3.5) 90CM HR37S (M)
Type of Device
SUTURE
Manufacturer (Section D)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, 08191
SP  08191
MDR Report Key7814203
MDR Text Key118229261
Report Number3003639970-2018-00511
Device Sequence Number1
Product Code GAM
Combination Product (y/n)N
PMA/PMN Number
K122734
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC0068956
Device Catalogue NumberC0068956
Device Lot Number118114
Was Device Available for Evaluation? No
Distributor Facility Aware Date07/25/2018
Initial Date Manufacturer Received 07/25/2018
Initial Date FDA Received08/24/2018
Supplement Dates Manufacturer Received11/07/2018
Supplement Dates FDA Received11/30/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-