• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US GII QUICK ANCHOR PLUS W/ #2 PANACRYL; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US GII QUICK ANCHOR PLUS W/ #2 PANACRYL; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE Back to Search Results
Catalog Number 212134
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement the complaint device is not being returned, therefore is unavailable for a physical evaluation.Furthermore, no lot numbers were supplied which precludes conducting a batch history review or a lot specific search in the complaints handling system.Therefore, we cannot determine what caused the user to experience the reported event, we cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: - incomplete.The lot number is currently unavailable; therefore, the expiration date is unavailable.This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event or Problem Description
It was reported by the affiliate in (b)(6) that during shoulder repair surgical procedure, it was observed that the suture on the gii quick anchor plus w/ #2 panacryl device broke while tying the knot.According to the report, the event occurred at the end of the surgery.It was reported that the surgeon used an arthrex anchor to complete the procedure.It was not reported if there were any delays in the surgical procedure.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII QUICK ANCHOR PLUS W/ #2 PANACRYL
Common Device Name
SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7814248
Report Number1221934-2018-52519
Device Sequence Number16777782
Product Code JDR
UDI-Device Identifier10886705001989
UDI-Public10886705001989
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K915889
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/20/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number212134
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/20/2014
Initial Report FDA Received Date08/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-