No radiographs or photographs have been provided to confirm the alleged event.It is unknown if the patient conformed to post-operative physical restrictions or if fusion was achieved.No further investigation can be completed at this time.Labeling review: ".Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s) loss of fixation, nonunion or delayed union." ".Correct selection of the implant is extremely important.The potential for success is increased by the selection of the proper size of the implant.While proper selection can minimize risks, the size and shape of human bones present limitations on the size and strength of implants.Metallic internal fixation devices cannot withstand the activity levels and/or loads equal to those placed on normal, healthy bone.These devices are not designed to withstand the unsupported stress of full weight or load bearing alone." ".These devices can break when subjected to the increased load associated with delayed union or nonunion.Internal fixation appliances are load-sharing devices that hold bony structures in alignment until healing occurs.If healing is delayed, or does not occur, the implant may eventually loosen, bend, or break.Loads on the device produced by load bearing and by the patient's activity level will dictate the longevity of the implant." ".Corrosion of the implant can occur.Implanting metals and alloys in the human body subjects them to a constantly changing environment of salts, acids, and alkalis, which can cause corrosion.Placing dissimilar metals in contact with each other can accelerate the corrosion process, which in turn, can enhance fatigue fractures of implants." ".Contraindications include but are not limited to: patients who are unwilling to restrict activities or follow medical advice.".
|
|
On (b)(6) 2010, patient underwent a transforaminal lumbar interbody fusion procedure.On (b)(6) 2017 approximately seven years post index surgery, patient was experiencing back pain.Radiographs revealed a pedicle screw fracture at s1 level and a facet screw fracture at the left contralateral side.On (b)(6) 2017 a revision procedure was performed due to adjacent segment disorder and fractured screw was replaced without any reported issues.
|