| Model Number |
85444 |
| Health Effect - Clinical Code |
No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On completion of the investigation a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
Stent was attempted to be deployed but balloon would not open.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis: based on the details of the complaint the stent was attempted to be deployed but the balloon would not open.Upon removal of the icast covered stent from the package the stent was loose in the package and no longer on the balloon.The balloon had been fully opened and one end of the stent was partially opened.The balloon was prepped per the instructions for use and the balloon was pressurized to 8atm.The balloon opened fully and there were no leaks detected.The stent when properly crimped on to the balloon leaves the impression of the stent frame on the balloon and indicates that the stent was properly crimped during manufacture.The stent frame crimp marks were clearly visible.The stent and balloon were very clean and had no bodily fluid stains on the catheter shaft or balloon.The flared end of the stent did have staining but was uneven.An evaluation of the balloon inflation lumens was conducted to ensure that both were patent as when the balloon is opened fluid is to be coming out of both the proximal and distal skive locations under the balloon.When the balloon was pressurized fluid could be seen entering the balloon at both skive locations.A thorough evaluation of the production lot history records was performed including the performance testing results of the product.Below is an overview of the quality and performance criteria that every lot of icast covered stent systems is tested to.This inspection requires that the catheter lot must pass the following: ability of the stent and delivery system to be passed through the labeled introducer sheath (6fr or 7fr).Ability to deploy the stent at nominal pressure (8atm).Ability to withdraw the deflated balloon catheter back through the labeled introducer sheath.Ability of the delivery system to withstand 10 inflate/deflate cycles at the rated burst pressure (12atm) without leaks or failures.Balloon burst testing.The balloon must burst over the rated burst pressure specified on the label (12atm).In addition to the final lot qualification testing there are multiple in-process inspections conducted that include the following: balloon hole skive dimensional verification.Stent securement testing.Proximal balloon weld tensile testing.Distal tip tensile testing.Catheter leak check.Atrium medical only release production lots that have passed the aforementioned performance and quality requirements.This production lot of icast covered stents passed all quality requirements.Conclusion: based on the proper deployment of the returned icast covered stent atrium medical cannot conclude that the device malfunctioned in any way.It is not known why the stent would not deploy during the case.
|
| |
|
Search Alerts/Recalls
|