COVIDIEN 760 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 760
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 08/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic has not received the suspect device/component from the customer for evaluation nor has the device been evaluated by the service engineer.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during use on a patient, the safety valve opened and a error message was generated.The patient was removed from the ventilator and placed on an alternate ventilator.
 
Additional Manufacturer Narrative
Correction to: evaluation codes (method, results, conclusion) device evaluation summary: a check of the log information confirmed that the measured value of the oxygen supply pressure was outside the specified value.The service engineer also evaluated the device and confirmed that the pressure transducer did not accurately measure pressure.The oxygen pressure transducer was replaced and the ventilator was returned to the customer after passing all testing.If the replaced part is returned for failure investigation, a supplemental medwatch report with the findings will be submitted once the investigation is complete.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
760 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
MDR Report Key7814465
Report Number8020893-2018-00396
Device Sequence Number1499832
Product Code CBK
UDI-Device Identifier10884521180918
UDI-Public10884521180918
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K990897
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number760
Device Catalogue NumberG-760110DIUU-JP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2018
Initial Date Received by Manufacturer 08/02/2018
Supplement Date Received by Manufacturer08/29/2018
Initial Report FDA Received Date08/24/2018
Supplement Report FDA Received Date01/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-