SMITH AND NEPHEW, INC. JOURN ART INS BCS STD 3-4 RT10; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 74023232
Health Effect - Clinical Code No Code Available (3191)
Date of Event 08/08/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was done on the patient.Reason for revision - pt slipped at work resulting in instability.At revision, insert exchanged for a journey bcs revision insert.At time of removal, unusual wear patten noticed and commented on by surgeon.Patient fit and healthy.
 
Additional Manufacturer Narrative
The associated complaint device was returned and evaluated.The lab analysis of the journ art ins bcs std 3-4 rt10 revealed the wear on the articular surface was likely due to normal articulation between the insert and femoral component.Burnishing was observed on the inferior surface of the insert.No discernible damage or material deformation was present on the insert post.No material or manufacturing deviations were observed during this investigation.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed part revealed no prior complaints for the listed batch.The clinical/medical team concluded, no clinical relevant documents were provided to conduct a thorough medical assessment.No medical assessment is warranted at this time.This complaint will be re-evaluated if more information becomes available.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURN ART INS BCS STD 3-4 RT10
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH AND NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7814914
Report Number1020279-2018-01625
Device Sequence Number1952716
Product Code JWH
UDI-Device Identifier03596010555007
UDI-Public03596010555007
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K042515
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/24/2022
Device Catalogue Number74023232
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 08/08/2018
Supplement Date Received by Manufacturer08/08/2018
Initial Report FDA Received Date08/24/2018
Supplement Report FDA Received Date01/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-