| Model Number |
MMT-7020A |
| Medical Device Problem Code |
Calibration Problem (2890)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that they experienced inaccurate readings and triggered threshold suspend.The customer's blood glucose was 186 mg/dl and sensor glucose was 60 mg/dl at the time of incident when it triggered threshold suspend.The sensor will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Inspected one random opened or used sensor and performed continuity resistance and sensor passed per specifications.Performed bicarbonate buffer test and sensor failed per specification due to high readings.See attached file for details.
|
| |
|