| Model Number |
MMT-7020A |
| Medical Device Problem Code |
Calibration Problem (2890)
|
| Health Effect - Clinical Code |
Blood Loss (2597)
|
| Date of Event |
03/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Inspected 1 opened/used sensor and performed continuity resistance test.Sensor passed per specification.Also, performed bicarbonate buffer test.Sensor passed per specifications with accurate readings.Also, performed visual inspection and checked insertion needle for burrs/hooks and dull needle tip defects and none was found.Insertion needle passed per specifications.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that they experienced inaccurate readings and triggered threshold suspend.The customer's blood glucose was 73 mg/dl and sensor glucose was 43 mg/dl at the time of incident when it triggered threshold suspend.The customer's blood glucose was 83 mg/dl at the time of call.The sensor was returned for analysis.
|
| |
|