| Catalog Number |
RLT281212 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
07/19/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The review of the manufacturing records verified that the lots involved in this event met all pre-release specifications.According to the gore® excluder® aaa endoprosthesis instructions for use (ifu), complications associated with the use of the gore® excluder® aaa endoprosthesis may include but are not limited to: endoleaks.
|
| |
|
Event or Problem Description
|
|
The following information was reported to gore: on (b)(6) 2018, this patient underwent emergency endovascular treatment for a ruptured abdominal aortic aneurysm and was treated with gore® excluder® aaa endoprostheses.On (b)(6) 2018, follow-up computed tomography imaging identified a type 1a endoleak resulting from the patient¿s aortic neck length with 10 mm only and therefore considered outside of the instructions for use.On (b)(6) 2018, a re-intervention was performed and it was unsuccessfully attempted to repair the endoleak by using aptus® staples.The situation was subsequently solved by injecting thrombin into the aneurysm sac.The patient tolerated the procedure.
|
| |
|