| Model Number |
MMT-7008B |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/26/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Inspected 1 opened/used sensor and performed continuity resistance test and sensor passed per specifications.Performed the sensor initialization test using a new lab transmitter with a insulin paradigm pump, connected the sensor to the transmitter and placed it in 100 bts solution, confirmed the communication icon was displayed and that the transmitter was flashing while connected to the sensor.Sensor passed per specifications.Cannot perform bts test as the sensor is beyond its expiration date.Also found sensor with cannula bent.Unable to confirm customer received sensor in said condition due to customer returned opened/used.A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that they experienced inaccurate readings and triggered threshold suspend.The customer's blood glucose was 189 mg/dl and sensor glucose was 40 mg/dl at the time of incident when it triggered threshold suspend.The sensor will be returned for analysis.
|
| |
|