During a sleeve gastrectomy, the subject device was used.In the procedure, the tissue pad of the subject device was peeled off.The intended procedure was completed with similar device.There was no patient injury reported.This is the report regarding the peeled off of the tissue pad.
|
|
This is a supplemental report to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.The lot number of the subject device is unknown.As a result of checking the manufacturing record for past six months from the delivery date, it was found no irregularities.Therefore, the exact cause of the reported event could not be conclusively determined.Based on the past similar cases, it was known that the tissue pad was damaged since the user continued to activate output without grasping tissue (including after the tissue already cut).The instruction manual of the device has already warned as follows; *do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.
|