MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC SPINAL VERTEBRAL BODY REPLACEMENT DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC SPINAL VERTEBRAL BODY REPLACEMENT DEVICE Back to Search Results
Catalog Number 2982001
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/31/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that intra-op, the inserter broke during the implantation.Implant was almost placed when one 'finger' of the insert er broke off.Final placement was completed with the help of compactor.No broken fragment of the instrument remained inside the patient.No patient complications were reported.
 
Additional Manufacturer Narrative
Product analysis: visual examination confirms one of the prong tips has broken off.The broken prong is missing and not returned for analysis.Fracture initiation appears to be located at the base of the prong.Optical examination of the fracture surface reveals a fairly quasi-brittle fracture, with no indication of fatigue, and a shear lip which suggests the direction of fracture.The above observations are consistent with bend stress overload.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPINAL VERTEBRAL BODY REPLACEMENT DEVICE
Common Device Name
SPINAL VERTEBRAL BODY REPLACEMENT DEVICE
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key7817846
Report Number1030489-2018-01169
Device Sequence Number16465369
Product Code MQP
UDI-Device Identifier00885074549736
UDI-Public00885074549736
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K062133
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2982001
Device Lot NumberPR11A019
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2018
Initial Date Received by Manufacturer 07/31/2018
Supplement Date Received by Manufacturer11/27/2018
Initial Report FDA Received Date08/27/2018
Supplement Report FDA Received Date12/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-