| Catalog Number |
07027737190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event occurred in (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer stated that they received an erroneous result for one patient sample tested with the elecsys prolactin assay on a cobas 8000 e 801 module.It was asked, but it is not known if the erroneous result was reported outside of the laboratory.The customer routinely tests samples for prolactin on their tosoh aia-2000 analyzer.The sample was initially tested on the aia-2000 analyzer, resulting with a prolactin value of 255.8 ng/ml.The physician questioned the result and asked for the sample to be repeated on the e 801 analyzer.The repeat prolactin result from the e 801 analyzer was 343.9 ng/ml.No adverse events were alleged to have occurred with the patient.The serial number of the e 801 analyzer was asked for, but not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The patient sample was treated with polyethylene glycol and repeated on the e 801 analyzer, resulting as 133.2 ng/ml (77% recovery).
|
| |
|
Additional Manufacturer Narrative
|
|
Both analyzers used for prolactin testing provided results above reference ranges.The investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|