During a cataract surgery the irrigation became clogged.After a while irrigation started again and a particle appeared in the anterior chamber.The particle came thru the sleeve.It was 1.8 mm x 0.5 mm in size, transparent, hard and plastic in appearance.The particle was removed from the patients eye with no reported impact.
|
|
Report of the lab analysis states that the white colored particle was microscopically examined and photographed using a camera-equipped optical stereo microscope.The white colored particle was analyzed using an energy-dispersive spectrometer (eds) equipped scanning electron microscope (sem).Eds uses the characteristic x-rays generated from a sample bombarded with electrons to identify the elemental constituents comprising the sample.This technique generates a spectrum of x-ray energy lines associated with specific elements.The eds is capable of qualitative and semi-quantitative chemical analysis of elements atomic numbers 6 through 94.The detected elements are normalized to total 100%.The nominal detection limit for most elements is 0.5%.Eds analysis indicated that the white colored particle consist primarily of silicon, chlorine, and aluminum.Lesser concentrations of sodium, carbon, oxygen, sulfur, calcium, potassium and magnesium were also detected.This is consistent with organic silicon, saline residue, and aluminum.The white colored particle was analyzed using a fourier transform infrared (ftir) spectrometer.Ftir is used in the characterization and identification of organic materials through the determination of the intramolecular bonds in the compound.Radiation in the ir region can be utilized by making use of the fact that specific frequencies are absorbed by specific interatomic bonds in organic molecules.Thus ir spectroscopy involves collecting absorption information and analyzing it in the form of a spectrum.The frequencies at which there are absorptions of ir radiation ("peaks" or "signals") can be correlated directly to intramolecular bonds within the material being analyzed.The ftir analysis of the white colored particle produced a spectrum consistent with that of silicone.The conclusion states that this analysis indicates that the white colored particle, consists primarily of silicone.Lesser concentrations of aluminum and saline residue were also detected.The lab report indicates that the particle is primarily silicone, which could have come from the silicone phaco sleeve, but the white color of the returned particle does not match the color of any of the sleeves we use.The silicone sleeve we put in bl5111 is blue; no white silicone is used in any of the components in bl5111.For this reason, we cannot say for certain that the returned particle came from our pack.No remedial, corrective, preventive or field safety correction action is required.Review of the complaint database revealed no other complaints for "fm/particulate" have been submitted against lot w0145.The lot history, trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.No corrective action is necessary the investigation is complete.
|