| Catalog Number |
PDP507G |
| Medical Device Problem Code |
Component Incompatible (1108)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/31/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown plastic surgery on (b)(6) 2018 and suture was used.During the procedure, it was noted that the suture looked thicker than usual.There were no adverse consequences to the patient.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).A top foil and a dispensed needle/suture combination were returned for analysis.During the visual inspection of sample, the swage and attachment area were as expected.The suture was examined along strand and an unusual color could be observed.Functional test was performed on the sample and the diameter met the requirements.No incorrect component was observed.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the sample received, no assembly - incorrect component was found.
|
| |
|
Search Alerts/Recalls
|