| Model Number |
EGIATRS45AMT |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, when the device was put in the patient¿s cavity during a laparoscopy-assisted distal gastrectomy, the jaw did not open, although the button was pressed many times, it did not open.The doctor appealed that all the buttons did not work at that moment.The cartridge was replaced with a new one.After that, the device could be used without problems.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection and functional testing confirmed there were no abnormalities that would have caused or contributed to the reported condition.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition; therefore, a review of the device history record was not performed.Our investigation was unable to determine a root cause or establish a relationship between the device and the reported condition.No enhancements or improvements were generated for the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|