| Model Number |
G21711 |
| Medical Device Problem Codes |
Dent in Material (2526); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Problem statement: the customer reported the following complaint issue: ¿before using this product for procedure, the side hole of this product pigtail was kink¿.Lab evaluation: 1 x zso-7-15 of lot # c1387439 was not returned to cirl for evaluation.Root cause: a definitive root cause for the customer complaint could not be determined as the device was not returned for evaluation.However, a possible cause of this complaint is that the kink occurred when straightening the stent.A precaution included in instructions for use, states "care must be exercised when straightening the pigtail curls* in order to avoid kinking or breaking the stent." as the device was not returned the complaint is confirmed based on the customers testimony.Documents review: a review of the manufacturing records for the device of lot # c1387439 did not reveal any discrepancy related to the complaint issue.There is no evidence to suggest that this issue affects the entire lot # c1387439; upon review of complaints this failure mode has not occurred previously with this lot # c1387439.Prd/fqc/pkg review: prior to distribution, all devices are subjected to 100% inspection to ensure device integrity.There is also a visual inspection of the product and packaging at packaging, packaging qc, and post sterile qc.Ifu review: it may be noted that according to the instructions for use, the user is instructed to: ¿visually inspect with particular attention to kinks, bends and breaks.If any abnormality is detected that would prohibit proper working condition, do not use¿.According to the instructions for use, the user is advised of the following: ¿care must be exercised when straightening the pigtail curls* in order to avoid kinking or breaking the stent".Summary: the complaint is confirmed based on customer testimony.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|
Event or Problem Description
|
|
Report is being submitted under the precedence of "stent kinked".Before using this product for procedure, the side hole of this product pigtail was kink.
|
| |
|
Additional Manufacturer Narrative
|
|
Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(6).Cook medical incorporated (cmi), 1025 acuff road, p.O box 4195, bloomington, indiana 47402-4195.Importer site establishment registration number: (b)(4).The customer reported the following complaint issue: ¿before using this product for procedure, the side hole of this product pigtail was kink¿.Lab evaluation: 1 x zso-7-15 of lot # c1387439 was returned to cirl for evaluation.A lab evaluation was held on 30 aug 2018.During the lab evaluation, the stent was found to be kinked at second and fifth porthole on tapered end.Wire guide stopped at second port hole.Root cause: a definitive root cause for the customer complaint could not be determined as the exact operational conditions of use could not be replicated in the laboratory setting.However, a possible cause of this complaint is that the physician may have kinked the device when removing it from the packaging or the device may have become kinked during transport or storage.The complaint is confirmed as the failure was verified in the laboratory and the stent was found to be kinked.Documents review: a review of the manufacturing records for the device of lot # c1387439 did not reveal any discrepancy related to the complaint issue.There is no evidence to suggest that this issue affects the entire lot # c1387439; upon review of complaints this failure mode has not occurred previously with this lot # c1387439.Prd/fqc/pkg review: prior to distribution, all devices are subjected to 100% inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.There is also a visual inspection of the product and packaging at packaging, packaging qc, and post sterile qc.Ifu review: it may be noted that according to the instructions for use, ifu0045-6, the user is instructed to: ¿visually inspect with particular attention to kinks, bends and breaks.If any abnormality is detected that would prohibit proper working condition, do not use¿.Summary: complaint is confirmed as the failure was verified in the laboratory.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|
Event or Problem Description
|
|
Report is being submitted under the precedence of "stent kinked".Before using this product for procedure, the side hole of this product pigtail was kink.
|
| |
|