Catalog Number DLF100310
Medical Device Problem Codes Material Frayed (1262); Stretched (1601)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Information regarding patient age, gender, weight, medical history, race, and ethnicity was not provided.Procode: krd/hcg.Physical manufacturer name: codman and shurtleff inc., dba depuy synthes products, inc.((b)(4)).Initial reporter - the customer contact information, including occupation, fax, and e-mail address was not reported.Physical manufacturer name: codman and shurtleff inc., dba depuy synthes products, inc.((b)(4)).The product is available for evaluation and testing; however, it has not been received to date (which is indicated as "other").Additional information will be submitted within 30 days of receipt.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.
 
Event or Problem Description
As reported by a healthcare professional, during a coil embolization of a moderately tortuous vessel, the physician started to deploy the 3mm x 10cm deltafill10 (dlf100310/s11155) thermo-mechanical coil when he noticed that it was frayed.The coil was removed from the sl-10 (stryker) microcatheter.No patient complications occurred as a result of the event.No additional intervention was required to prevent impairment.The product will be returned for analysis.No further information was provided.
 
Additional Manufacturer Narrative
(b)(4).As reported by a healthcare professional, during a coil embolization of a moderately tortuous vessel, the physician started to deploy the 3mm x 10cm deltafill10 (dlf100310/s11155) thermo-mechanical coil when he noticed that it was frayed.The coil was removed from the sl-10 (stryker) microcatheter.No patient complications occurred as a result of the event.No additional intervention was required to prevent impairment.No further information was provided.The 3mm x 10cm deltafill10 was returned for evaluation.The returned package label matched the complaint information.The introducer was observed to be frayed in to two strands and stuck in the v-notch of the resheathing tool.One strand of the introducer was observed to go through the proximal end of the resheathing tool while the other exited through the normal opening.The embolic coil was returned inside the green introducer.The device positioning unit (dpu) core wire was kinked at approximately 5 cm from the proximal end of the device.Using a microscope, the introducer was observed to be lodged in the v-notch of the resheathing tool.The introducer was removed from the v-notch and the v-notch was observed to be fractured.The embolic coil ball tip was seen present and intact.The distal outer sheath had not been softened, indicating that the detachment process had not been initiated.The proximal end of the embolic coil was seen to be stretched up to the socket ring.No other irregularities were seen on the embolic coil.Articulating joint was observed to be intact.A review of manufacturing documentation associated with this lot presented no issues during the manufacturing or inspection processes related to the reported complaint.The complaint of coil damage was confirmed; however, the exact circumstances surrounding the observed damage cannot be determined based on the condition of the returned device.The embolic coil was stretched at its proximal end.The complaint states that the device was removed from the microcatheter, so the stretching may have occurred sometime during retraction of the device because of excessive pull force.The frayed introducer indicates that the device was not properly unlocked.The instructions for use (ifu) instructs the following to properly unsheathe the device: ¿grasp the distal end of the re-sheathing tool between your left thumb and forefinger.Grasp the clear tab near the proximal end of the introducer sheath body with the thumb and forefinger of your other hand.Gently pull the clear tab of the introducer sheath out and away from the re-sheathing tool at a 45-degree angle to unlock the microcoil.Continue to pull the tab until an additional 0.5 to 1.0 in (1.3 to 2.5 cm) of the translucent material is exposed.Gently fold the translucent tab towards the distal end, and firmly grasp the distal end of the re-sheathing tool and the translucent tab between your thumb and forefinger¿.Pulling the tab straight out (in parallel to the dpu) will result in damage to the resheathing tool and may cause damage to the embolic coil.100% of devices are tested for sheathing functionality during manufacturing.Also, 100% of devices are inspected for irregularities to the embolic coil such as stretching.Thus, it is unlikely that the device left the manufacturing facility with the observed damage.The device did not present any obvious indication of manufacturing defect or anomaly that could have contributed to the event as reported.Neither the product analysis nor the device history review suggests that the failure could be related to the manufacturing process of the unit.Assignment of root cause for the event remains speculative and inconclusive, based on the minimal information provided and the evidence presented by the sample article; however, it is possible that procedural and handling factors, including device manipulation and the concomitant microcatheter, may have contributed to the reported failure and additional damages noted to the returned system.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective/preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).The purpose of this mdr submission is to report that the product has been returned for evaluation and testing; however, the engineering evaluation has not been completed.Additional information will be submitted within 30 days of receipt.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTAFILL10 3MM X 10CM
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
REFER TO SECTION H10
47709 fremont blvd
fremont CA 94538
MDR Report Key7818462
Report Number3008114965-2018-00725
Device Sequence Number19284677
Product Code KRD
UDI-Device Identifier10886704077039
UDI-Public10886704077039
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
K150319
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2019
Device Catalogue NumberDLF100310
Device Lot NumberS11155
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2018
Initial Date Received by Manufacturer 08/08/2018
Supplement Date Received by Manufacturer09/05/2018
10/05/2018
Initial Report FDA Received Date08/27/2018
Supplement Report FDA Received Date09/27/2018
10/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SL-10 (STRYKER) MICROCATHETER
-
-