| Model Number |
1570-02-135 |
| Medical Device Problem Codes |
Corroded (1131); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Calcium Deposits/Calcification (1758); Emotional Changes (1831); Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Muscular Rigidity (1968); Pain (1994); Loss of Range of Motion (2032); Tissue Damage (2104); Anxiety (2328); Distress (2329); Injury (2348); Deformity/ Disfigurement (2360); Depression (2361); Ambulation or Postural Difficulties (2544); Limited Mobility Of The Implanted Joint (2671); Test Result (2695); Not Applicable (3189); Metal Related Pathology (4530); Unspecified Tissue Injury (4559); Physical Asymmetry (4573); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
07/31/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Concomitant med products: asr uni femoral impl size 55; adapter sleeves 12/14 +2; asr acetabular cups 62; summit duofix tap sz7 std off.
|
| |
|
Event or Problem Description
|
|
Patient was revised to address suspected high metal ion levels, metallosis, and pseudotumor, also trunionosis documented.Surgeon removed depuy asr metal on metal hip and replaced it with modular pinnacle cup with ceramic on poly articulation.Doi: (b)(6) 2007; dor: (b)(6) 2018; right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Asr litgation received.In addition to what was previously alleged, litigation alleges injury, pain, disfigurement, and emotional distress.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).No device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
After review of medical records the patient was revised due to pain, stiffness, heterotopic ossification, adverse local tissue reaction, trunnionosis, psuedotumor, persistent limp, limited range of motion and gait dysfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Patient code: no code available (3191) used to capture the surgical intervention and medical device removal.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary = > no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary:no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Event or Problem Description
|
|
Pfs alleges inability to performed adl due to hip limitation, tension stress, leg length discrepancy, walking difficulty and sleep problem.Pfs also alleges that all injuries are continuing.In addition to what was previously reported in the medical records after review patient was revised to address walking difficulty.Doi: (b)(6) 2007.Dor: (b)(6) 2018.Right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: a4, b5, b6, b7, d4 (udi), d10, g2,, h6 (clinical, impact and medical device problem codes).H6 clinical codes: appropriate term/ code not available (e2402) is used to capture blood heavy metal increased and injury (e20).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: e1, e2, e3, h5.E3 initial reporter occupation: lawyer.
|
| |
|
Event or Problem Description
|
|
Pfs record received.In addition to what was previously alleged in the pfs, the patient had elevated inr, black staining tissues, mental and physical agonizing, depression, fear and anxiety.Post revision alleged hip pain, limited flexibility, limited range of motion, limited adl, leg strain, depression, anxiety and fear.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4) investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|