• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number R SERIES
Medical Device Problem Code Pacing Problem (1439)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event or Problem Description
Complainant alleged that while attempting to pace a patient (age & gender unknown), the device was unable to pace.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Additional Manufacturer Narrative
Zoll medical corporation evaluated the device.The reported malfunction could not be duplicated or confirmed.The device was put through extensive testing which included bench handling, performing 30 j tests, and pacer functionality stress testing using the returned multifunction cable without duplicating the reported malfunction.The device was recertified and returned to the customer.It is important to note that the electrode pads and ecg cable used during the reported event were not returned for evaluation.The clinical data was not available as part of the investigation and was the best opportunity for us to understand the circumstances related to the event.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key7819267
Report Number1220908-2018-02444
Device Sequence Number11214592
Product Code MKJ
UDI-Device Identifier00847946002374
UDI-Public00847946002374
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K060559/P160
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/06/2018
Supplement Date Received by Manufacturer08/06/2018
Initial Report FDA Received Date08/27/2018
Supplement Report FDA Received Date10/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-