| Model Number |
VS101505 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Occupation: health professional (hpro).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intra-operatively during a laparoscopic gastric bypass, multiple different 5mm instruments were introduced and caught in the middle portion of the cannula.This only happened on the first 4-5 passes through the cannula and then it no longer occurred.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection noted: that the trocar, cannula, circular and envelope seals appeared intact.The radially expandable sleeve was received.The obturator was received.A functional evaluation found that the device passed an air leak test.The cannula tip was checked with a 221.Pin gauge and was in spec.No resistance was noted.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|