| Model Number |
CA15L1 |
| Medical Device Problem Codes |
Break (1069); No Flow (2991)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a kidney tumor ablation, they noticed that the tube was torn and the water spilled.They held the two pieces of pipe together and finished the first ablation, but they could no longer perform another ablation.The generator had an alarm because of the leak and then turned off.They opened a new antenna to complete the case.They noted that pump also had a circulation issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary one device was received for evaluation.The returned product did not meet specification as received.Visual inspection found the luer was broken between the orange and clear inflow tubing.The reported condition was confirmed.The investigation found the tubing leaked at the orange/clear tubing junction.The connector was broken.Post market vigilance (pmv) could not determine how or why the connector broke.The investigation could not determine the root cause of the customer's report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant product/s: cagen1 cagen1 ablation generator x1 serial #:(b)(6); capump1 capump1 ablation pump x1 serial #:(b)(6); ca190rc1 ca190rc1 ablation reusable cable x1 lot #:ec16300897.H3 evaluation summary: one ca15l1 was returned for evaluation.Visual inspection noted the luer was broken between the orange and clear inflow tubing.The analysis found the tubing leaked at the orange/clear tubing junction and the connector was broken.The most likely cause was determined to be a weak joint from the supplier.Internal process improvements have been initiated to mitigate this issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|