DEPUY MITEK LLC US MILAGRO INTSCR 9X30MM *EA; ORTHOPAEDIC BONE SCREW, BIOABSORBABLE Back to Search Results
Catalog Number 231825
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/31/2014
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement: the complaint device is not being returned, it was discarded and therefore is unavailable for a physical evaluation.A batch record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed no other complaints of any kind for this lot of 765 devices that were released to distribution.We cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event or Problem Description
It was reported that as the surgeon was screwing the screw into the tibia, the tip of the implant broke.Report only, device was discarded by the customer.The procedure was completed with same like device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
ORTHOPAEDIC BONE SCREW, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7819950
Report Number1221934-2018-54549
Device Sequence Number1517352
Product Code MAI
UDI-Device Identifier10886705010806
UDI-Public10886705010806
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/31/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2015
Device Catalogue Number231825
Device Lot Number3638464
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/31/2014
Initial Report FDA Received Date08/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-