Model Number G35536
Medical Device Problem Code Material Perforation (2205)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, the advance 35 lp low profile balloon catheter was perforated.The defect was observed before starting the angioplasty procedure.There was no contact with the patient.According to the initial reporter, the patient did not experience any adverse effects nor require additional procedures due to this occurrence.Per the complainant, the complaint device will not be returned to the manufacturer for investigation.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation evaluation.A review of the complaint history, device history record, instructions for use (ifu), manufactures instructions, quality control, trends, and a visual inspection of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record revealed one other nonconforming event which could contribute to this failure mode.That nonconforming event was for failed leakage test however, this device was scrapped prior to order completion.It should also be noted there were no other reported complaints for this lot number.Additionally, though the device was not returned for evaluation, we did receive photos of the device.However, the images provided did not highlight the suspected failure.An attempt was made to obtain a representative device from the same lot but all devices of this lot had been distributed.Furthermore, an ifu is provided with the device, which states ¿the balloon is manufactured from an extra-thinwall, high-strength, minimally-compliant material.Particular care should be taken in handling the balloon to prevent damage.¿ it goes on to say ¿do not exceed rated burst pressure.Rupture of balloon may occur.Adhere to balloon inflation pressure parameters in the compliance card insert.Over-inflation may cause rupture of the balloon, with resultant damage to the vessel wall.Use of a pressure gauge is recommended to monitor inflation pressures." despite multiple attempts, we were unable to obtain any additional information.Without that additional information we were unable to determine if the failure of the device was attributed to manufacturing, device handling, user technique, and/or device failure.Measures are being conducted to address this failure mode.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Additional information was received from the distributor on (b)(6) 2019 stating the hole in the material was noted during the procedure and not prior to it as had been originally reported.While the balloon was in contact with the patient, an attempt was made to inflate it without success.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Additional information regarding the patient and/or event has been received since the previous medwatch report was sent.This information is detailed.
 
Event or Problem Description
Additional information regarding the patient and/or event has been received since the previous medwatch report was sent.This information is detailed in section h10.
 
Additional Manufacturer Narrative
Additional information was received from the customer.It was reported that the balloon was not inflated on the preparation table; it was only inflated in the target area of the patient.No holes were observed prior to the attempted usage of the device.Additional clarification regarding the information received has been requested from the customer.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 35 LP LOW PROFILE BALLOON CATHETER
Common Device Name
LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7819965
Report Number1820334-2018-02525
Device Sequence Number11349813
Product Code LIT
UDI-Device Identifier10827002355368
UDI-Public(01)10827002355368(17)200118(10)7600145
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K132020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other,use
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/18/2020
Device Model NumberG35536
Device Catalogue NumberPTA5-35-135-6-20.0
Device Lot Number7600145
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/16/2018
Supplement Date Received by Manufacturer10/16/2018
11/16/2018
02/28/2019
Initial Report FDA Received Date08/27/2018
Supplement Report FDA Received Date10/31/2018
11/19/2018
03/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-